| DS-NET TH8 | DS-NET BR4 | DS-NET V8 | DS-NET ACC2 | DS-NET AO4 | DS-NET DIO8 | |
|---|---|---|---|---|---|---|
| Input types | ||||||
| Voltage | ||||||
| Current | ||||||
| Resistance | ||||||
| Potentiometer | ||||||
| Pt100, Pt1000 | ||||||
| Thermocouple | ||||||
| Full, Half and Quarter Bridges | ||||||
| Frequency Signal | ||||||
| Pulse width | ||||||
| Counter | ||||||
| Time | ||||||
| Status | ||||||
| IPE/ICP Sensor | ||||||
| Output signal | ||||||
| Voltage | ||||||
| Current | ||||||
| Frequency Signal | ||||||
| Pulse Width | ||||||
| Status | ||||||
| Other data | ||||||
| Number of Channels | 8 | 4 | 8 | 2 | 4 | 8 |
| Data Rates (Hz) | 100 | 10 k | 100 k | 100 k | 100 k | 10 k |
| Standard connector | Screw terminal | Screw terminal | Screw terminal | Screw terminal | Screw terminal | Screw terminal |
| Optional connector | Thermocouple connector | DSUB 9 connector | BNC connector | DSUB 9, BNC connector | N/A | N/A |
DS NET documents |
||||||
Thermocouples and voltages in the range of ±80 mV Isolation voltage 100 VDC.
Modules can be replaced during measurement without pre-programming. Setup is loaded from the module's backplane.
Good thermal coupling by means of cold junction compensation per connector.
Optimized positioning of the interpolation points within the selected range, type B, E, J, K, L N, R, S, T, U.
24 bit ADU, 100 Hz sample rate per channel with 8 active channels, sum sample rate 800 Hz.
Digital filter, average, scaling, min/max storage, arithmetic, alarm.
Galvanic isolation of I/O-signals, power supply and interface Isolation voltage 500 VDC.
According to EN 61000-4 and EN 55011 standards.
Voltage, current, resistance, potentiometer, Pt100, Pt1000, thermocouples, measuring bridges.
24 bit ADU, 10 kHz sample rate per channel.
Good thermal coupling by means of cold junction compensation per connector.
16 virtual channels, linearization, digital filter, average, scaling, min/max storage, RMS, arithmetic, alarm.
Galvanic isolation of I/O-signals, power supply and interface Isolation voltage 500 VDC.
According to EN 61000-4 and EN 55011 standards.
Differential voltage, current via shunt connector.
Isolation voltage 100 VDC.
24 bit ADU, 10 kHz sample rate per channel with 8 active channels, sum sample rate 80 kHz.
Input: state, tare, memory reset.
Output: state alarm, threshold.
Linearization, digital filter, average, scaling, min/max storage, RMS, arithmetic, alarm.
Galvanic isolation of I/O-signals, power supply and interface Isolation voltage 500 VDC.
According to EN 61000-4 and EN 55011 standards.
Voltage, current, resistance, potentiometer, Pt100, Pt1000, thermocouples, measuring bridges, IEPE sensors.
24 bit ADU, 100 kHz sample rate per channel.
Input: state, tare, memory reset.
Output: state alarm, threshold.
16 virtual channels, linearization, digital filter, average, scaling, min/max storage, RMS, arithmetic, alarm.
Galvanic isolation of I/O-signals, power supply and interface Isolation voltage 500 VDC.
According to EN 61000-4 and EN 55011 standards.
Voltage ±10 V, current 4..20 mA selectable.
Isolation voltage 500 VDC.
100 kHz with 1 channel.
10 kHz with 4 channels.
Configurable as 2 counter, 2 frequency, or 2 PWM inputs, 1 frequency or PWM output, state in or output.
Frequency measurement up to 1 MHz (Chronos method).
Frequency output up to 10 kHz.
For/backward counter, quadrature counter with reference zero recognition (reset/enable), up to 1 MHz.
Measurement of duty cycle and frequency, output with variable frequency and/or duty cycle.
Galvanic isolation of I/O-signals, power supply and interface Isolation voltage 500 VDC.
According to EN 61000-4 and EN 55011 standards.
Configurable as counter, frequency, PWM and time inputs, frequency or PWM output, state in or output.
Process and host controlled.
Configurable as 2 counter, 2 frequency, or 2 PWM inputs, 1 frequency or PWM output, state in or output.
Frequency measurement up to 1 MHz (Chronos method).
Frequency output up to 10 kHz.
Up/down counter, quadrature counter with reference zero recognition (reset/enable), up to 1 MHz.
Measurement of duty cycle and frequency, output with variable frequency and/or duty cycle.
Galvanic isolation of I/O-signals, power supply and interface Isolation voltage 500 VDC.
According to EN 61000-4 and EN 55011 standards.